Error 2 on this model indicates that the imbalance sensor is tripped. If you run the unit with an empty rotor and it does not error, then it simply means that your load was imbalanced. Make sure to put vials into the unit opposite each other, as below. If it displays 'Error 2' with an empty rotor and you did not notice a vibration, then the imbalance sensor is broken and must be replaced. Kobelco sk210 service manual. Do not attempt to run the centrifuge without the imbalance sensor, as serious injury may result. I repair these machines, but you would have to ship it to me in Ohio, which would probably be very expensive.
Eppendorf 5415 R Centrifuge Repair Manual Ebook Eppendorf 5415 R Centrifuge Repair Manual currently available at www.cleopatralifehotel.com for review only, if you need complete ebook Eppendorf 5415 R Centrifuge Repair Manual please fill out registration form to access in our databases. Operating manual – Centrifuge 5427 R, 5.05 MB and for non-refrigerated centrifuges with imbalance system (5415 R, 5418 R. Centrifuge 5415 D / Centrifuge 5415 R – Eppendorf. Eppendorf micro test tubes Browse Eppendorf 5415d Service Manual – Minne HAHA. Eppendorf 5415d. Find the best prices on used Eppendorf instruments, equipment,.
Here is a rotor balance chart: Aug 13, 2015 . Dear Sir, In a 5402 Eppendorf Centrifuge, Error 1 states primarily for an EEPROM error. EEPROM is a tiny memory chip that stores individual details of the centrifuge, mainly the microprocessor time base (for correct time set operations), the mains voltage (115 or 230 V, to inform and set the motor current levels), and the imbalance level adjustment for the imbalance sensor. When the centrifuge powers up, these data are readen, and if all OK, the centrifuge is ready to go. If one of them corrupts, especially the 115 or 230 Volt information, we get error 1 and no start allowed. To check and reprogram this EEPROM we have to acess the service routine.
This is done by by jumping together pins 1 and 5 of X8 conector, on the base of the electronic board, located behind the power cable panel in the back of the centrifuge. There are about 20 setting or check parameters, but usually 1 or 2 of the described above need to be reprogrammed.
Free download program mastering c venugopal pdf writer. Please e-mail me for details: ph.